Nu Instruments | GC-C-IRMS Analysis with the Horizon IRMS and GC Interface

This application note demonstrates the performance of the Horizon IRMS coupled with the Nu GC-Interface for compound-specific carbon isotope analysis. Using a certified n-alkane standard, the study evaluates reproducibility, linearity and analytical accuracy over extended operation.

Typical IRMS chromatogram of B4 mix; collected ion current is plotted against time. A reference gas peak at the start of the run is used to calculate the δ13C value of each compound.

Gas chromatography-combustion-isotope ratio mass spectrometry (GC-C-IRMS) enables isotope analysis of individual compounds within complex mixtures. This application note explains how the Horizon IRMS integrates with an Agilent gas chromatograph through a dedicated GC-Interface designed for ease of operation and long reactor life.


To assess system performance, a certified Indiana University n-alkane standard mixture was analysed repeatedly over an eleven-hour period. The test evaluated the reproducibility and accuracy of δ¹³C measurements across compounds spanning a five-fold concentration range.


Results demonstrate excellent stability and analytical confidence. Standard deviations remained below 0.2‰ across all n-alkane compounds, with larger peaks achieving even greater reproducibility. The data also confirm excellent linearity and strong agreement with certified reference values, making the system well suited to demanding GC-C-IRMS applications.
 

Horizon IRMS Instrument

Key Takeaways:

  • Certified n-alkane testing demonstrates excellent δ¹³C measurement reproducibility.
  • Standard deviations below 0.2‰ confirm stable long-term performance.
  • Strong linearity and accuracy support reliable compound-specific isotope analysis.