Nu Instruments | Long-Term Stability of Dual-Mode Ion Counting Calibration 

This study evaluates the long-term stability of the Attom detector crossover calibration. Results demonstrate highly stable detector performance across varying signal intensities over an extended period.

Detector modes for the Attom HR-ICP-MS.


High resolution ICP-MS applications often require detection of both major and trace elements within a single analytical run. Achieving this demands a detector system with a wide dynamic range and stable calibration.


The Attom uses a unique attenuation system that extends pulse-counting performance without relying on analogue detector modes. Calibration data were collected over approximately one month.


Results showed that attenuation factors remained mass independent and stable to better than 1% RSD throughout the study period. This supports confident analysis of samples with widely varying concentrations.

Attom HR ICP-MS Instrument

Key Takeaways:

  • Stable crossover calibration maintained over one month.
  • Mass-independent attenuation performance demonstrated.
  • Supports wide dynamic range elemental and isotope analyses.